Search results for "living heart human model"
showing 2 items of 2 documents
Numerical simulation of transcatheter mitral valve replacement: The dynamic implication of LVOT obstruction in the valve-in-ring case.
2022
Transcatheter mitral valve replacement (TMVR) has been used for “off-label” treatment when annuloplasty band ring for mitral repair fails. However, the complex anatomy and function of the mitral valve may lead to fatal complications as a result of the left ventricular outflow tract (LVOT) obstruction in TMVR. We report the structural and hemodynamic response of LVOT obstruction resulting from TMVR with the Edwards SAPIEN 3 Ultra (S3) device. We modified the original Living Heart Human Model (LHHM) to account for a failed mitral valve with an annuloplasty band ring and simulated the cardiac beating condition in the setting of S3 device implantation. Findings demonstrated a high dynamic behav…
Patient-Specific Analysis of Ascending Thoracic Aortic Aneurysm with the Living Heart Human Model
2021
In ascending thoracic aortic aneurysms (ATAAs), aneurysm kinematics are driven by ventricular traction occurring every heartbeat, increasing the stress level of dilated aortic wall. Aortic elongation due to heart motion and aortic length are emerging as potential indicators of adverse events in ATAAs